MAIN MENU

MX1609KB

QuantumX termoelementos tipo K (16)

Amplificador para termopares de 16 canales de la
familia QuantumX;
Compatible con las siguientes tecnologías de
transductores:
- Termopares tipo K (NiCr-NiAl)
Características de cada canal:
- Velocidad de muestreo: máx. 600 S/s;
- Filtro paso bajo ajustable: Bessel, Butterworth;
- TEDS: identificación automática del punto de
medición (RFID);
- aislamiento galvánico;
- Toma: termo mini (verde);
Generalidades:
- 1 x Ethernet;
- 2 x FireWire;
- Tensión de alimentación: 10...30 VCC, máx. 6 W;
- Datos de calibración de fábrica según DIN/ISO
10012 guardados en el dispositivo (generar el
certificado con el MX Assistent)
Accesorio opcional:
- Paquete de 5 conectores macho termo mini con RFID:
1-THERMO-MINI.

img

QuantumX termoelementos tipo K (16)

expand_more
chevron_left
chevron_right

MX1609KB: High‑precision Temperature Measurements

The MX1609KB is a 16‑channel thermocouple amplifier designed for accurate and interference‑resistant temperature measurements. As part of the QuantumX family, it delivers exceptional precision, seamless system integration, and the flexibility needed for demanding R&D environments.

Key features

  • High‑precision temperature measurement with accuracy up to ±0.5 K
  • Fast response with sampling rates up to 600 Hz per channel for dynamic applications
  • 16 galvanically isolated channels with individually optimized cold junctions enabling exact measurement results
  • Advanced signal conditioning, including Bessel and Butterworth low‑pass filters
  • TEDS support for automatic sensor identification
  • Robust connectivity via Ethernet (PTP) and FireWire for easy system integration
  • Factory calibration data stored on the device according to ISO10012

Real-time data acquisition of temperature measurements

The MX1609KB is optimized for type K thermocouples, covering a wide temperature (linearization) range from –100 to +1300 °C/–148 to +2372 °F. With internal cold junctions for each input, you are ensured maximum accuracy, while active low‑pass filtering provides clean, interference‑free signals. With up to 600 Hz per channel, the module captures rapid temperature changes with clarity, making it ideal for dynamic thermal analysis.

Typical applications

  • Highly dynamic temperature measurements in research and development
  • Climate chamber testing and complex thermal modeling
  • Multidimensional temperature mapping using 16 individually scalable channels
  • Fast‑changing thermal processes requiring high sampling rates and low noise

Fast, easy integration

All applications benefit from the module’s precision, electrical isolation, and seamless integration into larger QuantumX systems. And due to its wide range of connection options, Ethernet, FireWire, PROFINET and EtherCAT, you are sure of optimal connectivity and effortless system integration. But it is the integrated TEDS capability, where the module automatically identifies connected sensors, that allows you to streamline setup and minimize configuration time.

Comparable modules

The MX1609 series offers top quality with an outstanding price-to-performance ratio. For comparison, the MX1609TB version supports type T thermocouples,

For more challenging tasks in an electrical environment, the QuantumX MX809B insulated thermocouple amplifier is available.

Stay connected with the latest firmware and software

All QuantumX modules are delivered with the latest firmware. For existing modules, it is easy to update the firmware by simply downloading the latest firmware package.

HBK also provides both native and third-party software packages and drivers allowing you to fully leverage the QuantumX ‘open’ data acquisition infrastructure for integration with a great number of operating, analysis and automation software. This includes:

  • MX Assistant: a modern and free device or system assistant that supports all module functions
  • catman®EASY / AP / Enterprise: The standard DAQ software used with the QuantumX platform
  • Third-party drivers: For example, LabVIEW™, Visual Studio® .NET and CANape®

Go to our download site for more information and a complete list of options.

Did you know? You can also connect Tescia data acquisition software to your QuantumX module via the MX Assistant utility. Talk to one of our experts to find out more.

Idioma keyboard_arrow_down
MX1609KB - Hoja de cracterísticas Hoja de características English
MX410 - Guía rápida Guía rápida English, Italian, French, Portuguese, Chinese, Korean, German, Spanish, Japanese
QuantumX - Manual de empleo Manual de empleo English
MX1609KB - Hoja de cracterísticas Hoja de características Chinese
MX1609KB - Hoja de cracterísticas Hoja de características Japanese
MX1609KB - Hoja de características Hoja de características German
QuantumX - Manual de empleo Manual de empleo Japanese
QuantumX - Manual de empleo Manual de empleo German
QuantumX - Manual de empleo Manual de empleo Chinese

Sales documents

QuantumX - Folleto Folleto English
QuantumX - Folleto Folleto German
QuantumX - Folleto Folleto Portuguese
QuantumX - Folleto Folleto Chinese
QuantumX - Folleto Folleto Japanese

Certificates and Approvals

QuantumX - Declaración de Conformidad Declaración de Conformidad English
MX1609KB - CAD CAD Step Files
MX1609KB-R - CAD CAD Step Files
Key attributes: DAQ Systems
Número de entradas 16
Entrada Thermocouples
Señal de salida None
Interfaz TCP/IP, Firewire
Clase de precisión 0.7
Grado de protección IP20
Frecuencia de muestreo 500 Hz
Carcasa Standalone
Capacidades de tiempo real None