MX809B
QuantumX family 8-channel measurement module for
isolated temperature measurements with thermocouples
or for measuring el. voltages;
Supports the following transducer technologies:
- Thermocouple types K, J, T, B, E, N, R, S, C with
internal cold junction for each channel;
- El. voltage: 5 V;
The following applies for each channel:
- Electrically isolated in accordance with EN60664:
1000 V, transient-proof up to Ut 2500 V (against
other channels, electrical supply, communication,
case);
- Measurement categories (safety): CAT II 600 V, CAT
III 300 V;
- Data rate: max. 600 S/s;
- 24-bit ADC;
- Digital low-pass filter: Bessel, Butterworth,
linear phase;
- Socket: Thermo-Mini (male, copper contact) prepared
for HBM insulating cap;
General:
- For temperature measurements on components that
conduct high voltage;
- Suitable for mobile and stationary applications
with electrical components;
General connections:
- 1 x Ethernet with PTP support;
- 2 x FireWire;
- Voltage supply: DC 10...30 V, max. 6 W;
- Factory calibration data in accordance with
DIN/ISO 10012 saved on the device (generate
certificate using MX Assistent)
Optional accessories:
- 1-ITC-K1000: thermocouple;
- 1-CON-A1018: package with 4 insulating caps for
Thermo Mini;
- 1-CON-S1016: Package with 4 type K couplings;
- 1-CON-S1017: package with 4 copper-copper couplings.
The new QuantumX MX809B module provides eight channels for the measurement of temperatures and voltages at a high potential. Every channel supports all customary thermocouples such as type K, J, N, T, E, and many others as well as, in general, voltages in the range of +/- 5 V. The module offers an innovative insulation concept to meet all safety requirements: Double-insulated measuring leads and standard thermo Mini connectors (IEC, ANSI) as well as a patented contact protection and safety concept (insulating cap) prevent direct contact with hazardous live parts to ensure highly reliable measurement.
Typical fields of applications for the MX809B include the testing and analysis of electric motors, generators, rectifiers or inverters, chargers, and, in particular, energy storage systems (high-voltage batteries) and fuel cells. In battery tests, the amplifier allows for precise measurement of minimal cell voltages and temperatures at a potential of up to 1,000 V. Safety is guaranteed even with voltage peaks of up to 2,400 V. Other modules of the QuantumX series can be added as desired. The complete system is ideally suited for use in the field and in the laboratory.
To make measurement results reliable and dependable, the calibration of transducers, amplifiers, or complete measurement chains is required. For QuantumX, the working standard calibration certificate according to ISO 10012 is stored in the module. The DAkkS certificate (ISO 17025) from HBM is available on request.
Unique coupling concept preventing direct contact with hazardous live parts to ensure highly reliable measurement
Flexibility to meet varying requirements thanks to the possibility of defining the sensor’s cable length and implementing the wiring as required in advance
Both centralized and distributedmodules allow for easy and individual integration
| QuantumX MX809B - Data Sheet | Data Sheets | English |
| MX809B - Data Sheets | Data Sheets | Japanese |
| MX809B - Data Sheets | Data Sheets | German |
| MX809B - Data Sheets | Data Sheets | French |
| MX809B - Data Sheets | Data Sheets | French |
| MX410 - Quick Start Guide | Quick Start Guide | English, Italian, French, Portuguese, Chinese, Korean, German, Spanish, Japanese |
| MX403B - Operating Manuals | Manual | English, French, German |
Sales documents
| QuantumX - Brochure | Brochure | English |
| QuantumX - Brochures | Brochure | German |
| QuantumX - Brochures | Brochure | Portuguese |
| QuantumX - Brochures | Brochure | Chinese |
| QuantumX - Brochure | Brochure | Japanese |
Certificates and Approvals
| QuantumX MX809B - EU Declaration of Conformity | Declaration of Conformity | English |
| MX809B - CAD | CAD Step Files |
| Number of inputs | 8 |
| Input | Thermocouples, Electrical voltage |
| Output | None |
| Interface | TCP/IP, Firewire |
| Accuracy class | 0.02 |
| Degree of protection | IP20 |
| Sample rate | 500 Hz |
| Housing | Standalone |
| Real-time capabilities | None |