MAIN MENU

MX840-PAKEASY

QuantumX paquete: MX840B + catmanEASY

Paquete de producto formado por:
Amplificador universal de 8 canales de la familia
QuantumX;
Amplificador universal de 8 canales de la familia
QuantumX;
Compatible con las siguientes tecnologías de
transductores (total de 8 canales):
- 8 x galgas extensométricas, 1/2 puente y puente
completo (CC o frecuencia portadora de 4,8 kHz);
- 8 x tensión (100 mV, 10 V, 60 V);
- 8 x transductores piezoeléctricos alimentados por
corriente (IEPE, ICP);
- 8 x piezorresistivos, puente completo;
- 8 x termopares (tipos K, N, R, S, T, B, E, J, C);
- 8 x termómetros de resistencia (Pt100, Pt1000);
- 8 x resistencia (5000 ohm);
- 8 x transductores potenciométricos;
- 8 x inductivos, 1/2 puente o puente completo, LVDT;
- 8 x corriente (±20 mA);
- 4 x codificador rotativo (con/sin indexación),
contador, frecuencia (rpm), protocolo SSI;
- 1 x bus CAN (recepción de 128 señales, transmisión
de 7 canales DAQ);
Características de cada canal:
- Velocidad de muestreo: máx. 40 kS/s (ancho de banda:
7 kHz);
- Filtro paso bajo ajustable (Bessel, Butterworth);
- Compatibilidad con TEDS: identificación automática
de transductores;
- Alimentación del sensor: 5...24 V, 0,7 W
(módulo: 2 W);
- Todas las entradas están eléctricamente aisladas;
- Toma: DSubHD 15 polos;
Generalidades:
- 1 x Ethernet (PTP), 2 x FireWire;
- Tensión de alimentación: 10...30 VCC, máx. 13 W;
- Datos de calibración de fábrica según DIN/ISO 10012
guardados en el dispositivo (generar el certificado
con el MX Assistent)
El paquete de productos tiene un descuento del 10% e
incluye además:
- Fuente de alimentación (1-NTX001)
- 8 conectores macho para transductores con TEDS
(1-SUBHD15-MALE)
- Cable Ethernet cruzado (1-KAB239-2)
- Software catman®Easy de HBM, incluido mantenimiento
de software durante los primeros 12 meses
Accesorios disponibles:
- Conexión para termopares: 1-THERMO-MXBOARD;
- Conexión para galga extensométrica en circuito 1/4
de puente: 1-SCM-SG120/350;
- Conexión de 10 V o IEPE a través de BNC:
1-SUBHD15-BNC;
- Conexión de 300 V CAT II: 1-SCM-HV.

img

QuantumX paquete: MX840B + catmanEASY

expand_more
chevron_left
chevron_right

QuantumX MX840B: 24-Bit Universal Measuring Amplifier for All Common Sensor Technologies

The MX840B modules are the universal measuring amplifiers of the QuantumX series. This means that sensors and transducers based on 17 different sensor technologies can be connected to each of the 8 channels available on these amplifiers (see complete survey below). Owing to the innovative “Plug & Measure” technology, the QuantumX module automatically identifies your transducer and is immediately ready to start measuring. 

The universal measuring amplifiers of the QuantumX family thus substantially minimize the setup time in the test stand and are the ideal choice for frequently changing measuring tasks. As always with the QuantumX, the universal measuring amplifiers can be employed as stand-alone instruments or combined with other modules; this can be achieved in various topologies (centralized, distributed, hybrid) as well as by connecting them to a PC (e.g. with HBM catman software) or a data recorder (e.g. for mobile measurements). The resulting flexibility and scalability are definitely unique in the market.

The universal modules provide a 24-bit analog-to-digital converter, sample rates of 40 kS/s per channel, and active low-pass filters and are renowned for their high precision.

The combination of precision, flexibility, and measurement setup at lightning speed makes each QuantumX universal measuring amplifier an “all-purpose tool” for demanding and highly efficient data acquisition in measurement, which has already been used in tens of thousands of applications. Typical applications include test stands, service tasks, mobile data acquisition, and monitoring. 

  • Accuracy class: up to 0.05%
  • Sample rate per channel: up to 40 kS/s

catman®Easy

catmanEasy is the basic version of the catman Data Acquisition Software which includes online and offline setup, online visualizations, and online maths as well as recording functionalities. catman Easy can be combined with two different add-on modules:

  • EasyMath: Mathematics module and auto sequences for catman Easy.
  • EasyScript: Free programming under VBA in measurement and analysis mode.

Time-saving and efficient

  • Connection of 17 different transducer technologies per channel
  • Plug & Measure: Automatic sensor identification and channel configuration
  • Fast calibration of transducers or validation through internal shunt connection
  • Custom configuration of individual channels
  • Wide ambient temperature range (-20 °C to +65 °C) and robust design allow use in any location
  • Distributed system and short, pre-wired sense leads

Minimized setup time in the test stand and an ideal fit for frequently changing measurement tasks, due to immediate sensor identification and universal connectors.

Precise and reliable

  • 24-bit analog-to-digital converter per channel
  • Individual sample rates of up to 40 kS/s
  • Active low-pass filter
  • Accuracy classes ranging from 0.05 to 0.1 (depending on the sensor technology)
  • Working standard calibration certificate according to ISO 10012 stored in the module
  • DAkkS certificate (ISO 17025) from HBM on request

Reliable results and proven high quality from HBM ensure successful tests at any time!

Compact, mobile, and scalable

  • The modules can be used as stand-alone instruments or combined with other measuring modules
  • Many measuring modules for special tasks are available (e.g. high-precision, torque, full bridges, etc.)
  • Different topologies (centralized, distributed, hybrid)
  • Real-time integration with the Ethernet fieldbus (CX27B), CAN (MX471B), or voltage output (MX878B)
  • Open software concept (Visual Studio .NET, LabVIEW, etc.)

From one to thousands of channels: Measuring equipment that is scalable to fit your requirements and integrates with your system.

Idioma keyboard_arrow_down
MX840B - Hoja de cracterísticas Data Sheets English
catman - Hoja de cracterísticas Data Sheets English
MX840B - Hoja de cracterísticas Data Sheets German
MX840B - Hoja de cracterísticas Data Sheets Chinese
MX840B - Hoja de cracterísticas Data Sheets Japanese
catman - Hoja de cracterísticas Data Sheets French
catman - Hoja de cracterísticas Data Sheets German
MX410 - Guía rápida Quick Start Guide English, Italian, French, Portuguese, Chinese, Korean, German, Spanish, Japanese
catman - Manual de empleo Manual English
QuantumX - Manual de empleo Manual English
catman - Manual de empleo Manual German
QuantumX - Manual de empleo Manual German
QuantumX - Manual de empleo Manual Japanese
QuantumX - Manual de empleo Manual Chinese

Sales documents

QuantumX - Folleto Folleto English
QuantumX - Folleto Folleto German
QuantumX - Folleto Folleto Portuguese
QuantumX - Folleto Folleto Chinese
QuantumX - Folleto Folleto Japanese

Certificates and Approvals

QuantumX - Declaración de Conformidad Declaration of Conformity English
MX840B - CAD CAD Step Files
Key attributes: DAQ Systems
Número de entradas 8
Entrada Strain gauge full/half bridge, Electrical voltage, Standardized current (4-20mA), Inductive bridge, LVDT, SSI, IEPE / CCLD, Piezo resistive bridge, Thermocouples, RTD (Pt, Ni), Resistance (PTC, NTC, KTY), Digital encoder, Timer digital input, CAN
Señal de salida CAN bus
Interfaz TCP/IP, Firewire
Clase de precisión 0.01, 0.05, 0.1
Grado de protección IP20
Frecuencia de muestreo 40000 Hz
Carcasa Standalone
Capacidades de tiempo real None