MAIN MENU

P111I-4

Sonda remota, corrente 4 canais

Sonda remota corrente 4 canais
isolada mediante link de fibra óptica
*** requisitos ***
- quatro canais de entrada de corrente analógicos
- entradas diferenciais balanceadas, isoladas,
com duas resistências de carga de precisão
0.33 Ω / 0.1 Ω
- conector padrão SubD de 9 pinos para transdutores
de corrente
- seis faixa de entrada de corrente até ±2 A
- entrada de voltagem adicional ±50 mV até ±20 V
para pinças de corrente com conector BNC separado
- isolamento 60 V para evitar loops de terra
- precisão de corrente (banda larga, pior cenário)
0,05% da leitura, 0,005% da faixa ± 200 μA
- precisão de potência, quando igualada com sonda
remota de voltagem 0,015% da leitura, 0,02% da
faixa
*** requisitos ***
placa receptora GN800B

img

Sonda remota, corrente 4 canais

expand_more
chevron_left
chevron_right

Current Remote Probe 

The P111I-4 Current Remote Probe is the current counterpart to the voltage probe. The device provides four current inputs up to ±2 A for direct connection to current transducers (CTs) and a sample rate of up to 2 MS/s. It also offers voltage inputs for current clamps or Rogowski coils.

  • Current accuracy of 0.025% reading +0.01% range
  • Bandwith of 2.3 MHz (-3dB)
  • Single fiber connection back to GN800B power analyzer card
  • 60 V isolation to prevent ground loops
  •  Standard 9 pin D-Sub connector for current transducer or voltage input for current clamps with separate BNC connector

Distributable

Can be used as part of a distributed measurement system where modules are situated close to the measurement object with cable length up to 100 metres. Ideal for large scale measurements.

 

Minimise EMC issues

Receiver card and mainframe separate for closer positioning of card to measurement point, reducing EMC interference. Glass fibre cables enable EMC-free data transfer to mainframe, ensuring EMC immunity through fibre-optic connections.

 

Safe and simple cabling – less is more

Our decentralised receiver card reduces cabling needs significantly, while the use of glass fibre cables ensures rapid and EMC-free data transfer capabilities of up to 100 metres.