See All Acoustic End-of-Line Test Systems See All DAQ and instruments See All Electroacoustics See All Software See All Transducers See All Vibration Testing Equipment See All Academy See All Resource Center See All Applications See All Industries See All Insights See All Services See All Support See All Our Business See All Our History See All Our Sustainability Commitment See All Global Presence
arrow_back_ios

Main Menu

See All Actuators See All Combustion Engines See All Durability See All eDrive See All Production Testing Sensors See All Transmission & Gearboxes See All Turbo Charger See All DAQ Systems See All High Precision and Calibration Systems See All Industrial electronics See All Power Analyser See All S&V Hand-held devices See All S&V Signal conditioner See All Accessories See All DAQ Software See All Drivers & API See All nCode - Durability and Fatigue Analysis See All ReliaSoft - Reliability Analysis and Management See All Test Data Management See All Utility See All Vibration Control See All Acoustic See All Current / voltage See All Displacement See All Load Cells See All Pressure See All Strain Gauges See All Torque See All Vibration See All LDS Shaker Systems See All Power Amplifiers See All Vibration Controllers See All Accessories for Vibration Testing Equipment See All Training Courses See All Articles See All Whitepapers See All Acoustics See All Asset & Process Monitoring See All Custom Sensors See All Data Acquisition & Analysis See All Durability & Fatigue See All Electric Power Testing See All NVH See All Reliability See All Smart Sensors See All Vibration See All Weighing See All Automotive & Ground Transportation See All Calibration See All Installation, Maintenance & Repair See All Support Brüel & Kjær See All Release Notes See All Compliance See All Our People
arrow_back_ios

Main Menu

See All CANHEAD See All GenHS See All LAN-XI See All MGCplus See All Optical Interrogators See All QuantumX See All SomatXR See All Accessories See All Accessories See All BK Connect / Pulse See All API See All Microphone Sets See All Microphone Cartridges See All Acoustic Calibrators See All Special Microphones See All Microphone Pre-amplifiers See All Sound Sources See All Accessories for acoustic transducers See All Experimental testing See All Transducer Manufacturing (OEM) See All Accessories See All Non-rotating (calibration) See All Rotating See All CCLD (IEPE) accelerometers See All Charge Accelerometers See All Impulse hammers / impedance heads See All Cables See All Accessories See All Electroacoustics See All Noise Source Identification See All Environmental Noise See All Sound Power and Sound Pressure See All Noise Certification See All Industrial Process Control See All Structural Health Monitoring See All Electrical Devices Testing See All Electrical Systems Testing See All Grid Testing See All High-Voltage Testing See All Vibration Testing with Electrodynamic Shakers See All Structural Dynamics See All Machine Analysis and Diagnostics See All Process Weighing See All Calibration Services for Transducers See All Calibration Services for Handheld Instruments See All Calibration Services for Instruments & DAQ See All On-Site Calibration See All Resources See All Software License Management

Structural Testing Part One deals with mechanical mobility measurements are an important stage of the engineering design process for dynamic mechanical structures. Determination of the dynamic characteristics of structures can be used to improve and optimize the structural performance. 

Optimizing the Structural Performance

Mechanical mobility measurements are an important stage of the engineering design process for dynamic mechanical structures. Determination of the dynamic characteristics of structures, including noise and vibration, can be used to improve and optimize the structural performance. 

The main dynamic properties of a mechanical system can be defined by means of the Frequency Response Function (FRF) as the ratio of measured excitation force applied to the structure and the corresponding motion of the structure.

The motion of a structure can be described in terms of displacement, velocity or acceleration. The corresponding Frequency Response Functions are compliance, mobility and accelerance. In a general sense the term ‘mobility measurement’ is used to describe any form of FRF. For modelling, the FRF most commonly used is compliance. The FRF generally used for measurements is accelerance, since the most convenient motion transducer is the accelerometer. Compliance, mobility and accelerance are algebraically related, the measurement of any one of them can be used for calculation of the others.

knowledge, resource center, articles, paconsult

Excitation Waveform in Mobility Measurements

For mobility measurements the structure must be excited by a measurable dynamic force, but there is no theoretical restriction as to waveform, or to how the excitation is implemented. 

In this discussion we will limit ourselves to waveforms which have energy distributed over a wide band of frequencies. These can simultaneously excite the structure over the entire frequency range of interest. Certain parameters should be considered before choosing the excitation waveform:

  • Application
  • Spectrum control
  • Crest factor
  • Linear/non-linear structure
  • Speed of test
  • Equipment available 


If the purpose of the test is only to measure natural frequencies, then the precision required is much less than when the measurements are to form the basis for a mathematical model. The cost of extra precision lies in the time taken for the measurements, and in the cost of the instrumentation. 

Spectrum control is the capacity to limit the excitation to the frequency range of interest. The dynamic range of an FRF is often very large, when measured between the highest resonance peak and the deepest antiresonance. Since the excitation waveform is generally chosen to have an ideally flat spectrum, it follows that the response spectrum will have the same large dynamic range as the FRF. If the structure is excited only in the frequency range of interest, the dynamic range of the measurement is minimized. This results in a better signal-to-noise ratio, and cleaner data.

The crest factor describes the ‘peakiness’ of the signal. It is defined as the ratio between the peak and the standard deviation (RMS) in the signal. A high crest factor in the excitation waveform has two disadvantages:

  • The signal-to-noise ratio is decreased, since the instrumentation must make allowances for the peaks, and some of the signal is lost in the existing noise
  • High peak forces may provoke non-linear behaviour in the structure


An expectation of non-linear behaviour in a structure raises the question: ‘Do we want to describe the non-linear behaviour, or do we want to make a linear approximation?

Modal analysis assumes linear systems and uses linear models. If we deal with a structure exhibiting some nonlinear behaviour, we generally attempt to make the best linear approximation. Selecting a waveform that excites the structure over a wide variety of levels randomizes the nonlinear behaviour which is then averaged out. To study nonlinearities, sinusoidal excitation with maximum amplitude control is generally used.