Connection between the sensor and the amplifier
Connection between the sensor and the amplifier
Only the 4-wire circuit, or HBM's patented Kreuzer circuit, enables different cable resistances to be compensated for. A known electric current flows through the resistor via two of the leads. The voltage drop at resistor RKab1 is corrected (at high impedance) via two additional leads.
The Kreuzer circuit measures the voltage across resistor RKab2 and adds it to the excitation. The voltage and thus the current through completion resistor Rerg are independent of the cable resistance. Zero point and sensitivity errors resulting from cable effects are electronically compensated for.
https://www.hbm.com/en/3458/tips-and-tricks-successfully-compensating-for-lead-resistances/
Only the 4-wire circuit, or HBM's patented Kreuzer circuit, enables different cable resistances to be compensated for. A known electric current flows through the resistor via two of the leads. The voltage drop at resistor RKab1 is corrected (at high impedance) via two additional leads.
The Kreuzer circuit measures the voltage across resistor RKab2 and adds it to the excitation. The voltage and thus the current through completion resistor Rerg are independent of the cable resistance. Zero point and sensitivity errors resulting from cable effects are electronically compensated for.
https://www.hbm.com/en/3458/tips-and-tricks-successfully-compensating-for-lead-resistances/
1 Thermally strip 5mm of insulation off the wire to be attached to the strain gauge.
Thermal stripping prevents damage which can occur from mechanical stripping with pliers.
1 Thermally strip 5mm of insulation off the wire to be attached to the strain gauge.
Thermal stripping prevents damage which can occur from mechanical stripping with pliers.
Disadvantage: Interference (caused by electric or magnetic fields as well as thermoelectric and galvanic voltages in the measuring circuit) is fully amplified.
Disadvantage: Interference (caused by electric or magnetic fields as well as thermoelectric and galvanic voltages in the measuring circuit) is fully amplified.
Disadvantage: Limited bandwidth
Disadvantage: Limited bandwidth