A sensor's exact
nominal (rated) sensitivity can be found in the test certificate. In most cases, this sensitivity is
2 mV/V, with the nominal (rated) force being applied.
As already mentioned above, the amplifier systems feed the measuring bridge with a bridge excitation voltage. This bridge
excitation voltage often is 5 V. If your force transducer's nominal (rated) sensitivity is 2 mV/V,
10 mV are available at the amplifier's
input stage when the force transducer is loaded with nominal (rated) force. For example, when you use an
S2M/100 N (100 N nominal (rated) force, 2 mV/V sensitivity at nominal (rated) force) and load it with 100 N you obtain 10 mV.
In general, the measured value is not always 100 N; in fact, it should also be possible to detect
smaller values. If you want to have a measurement signal resolution of
0.1 N,
10 µV are available at the
input stage.
If your amplifier has a
resolution of 100,000 divisions, which is still less than a high-precision instrument such as
HBM's DMP41 offers, this corresponds to the relationship between the height of the Eiffel tower (321 m) and the thickness of a CD case.
Requirements are significantly higher with increasingly complex measurement tasks.
The better the connection corresponds to the measurement task the more reliable are aour measurement results.