Type 4944 is a ¼-inch Prepolarized Pressure-field Microphone designed for high-level and high-frequency measurements over a frequency range of 4 Hz To 70 kHz.

img

TYPE 4944

expand_more
chevron_left
chevron_right

Prepolarized ¼″ Pressure-field Microphone Type 4944 is specially designed for high-level and high-frequency measurements. By using new materials and a new way of mounting the diaphragm, the microphone is geared to withstand rough handling. The assembly of the microphone in a clean room environment ensures that the microphone can be used in high-humidity environments and still produce reliable results. Type 4944 can be used with sound level meters to measure high sound pressure levels

  • High-level measurements
  • High-frequency measurements
  • Flush mounting
  • In small closed couplers, close to hard reflective surfaces or flush-mounted
  • The sensitivity has been optimised to allow measurements of high sound pressure levels without clipping in the built-in CCLD preamplifier.
TEDS Microphones Data Sheets English
Prepolarized Pressure-field 1/4'' Microphone Data Sheets English
TYPE 4944
Standard IEC 61094
Sound field Pressure field
Polarization voltage 0V
Sensitivity 0.9 mV/Pa
Lower limiting frequency 4 Hz
Higher limiting frequency 70 kHz
Inherent noise 30 dB(A)
Upper limiting of dynamic range 170 dB
Diameter 1/4 inch
Venting Side
Minimum Temperature -20 °C
Maximum Temperature 100 °C
Minimum Temperature -4 °F
Maximum Temperature 212 °F

DESIGN AND ROBUSTNESS

The shape of the microphone front ensures excellent microphone performance when flush-mounted. The laser-welded diaphragm on the microphone housing ensures that the sensitivity is resistant to rough handling during flush mounting.

Calibration

The sensitivity can be calibrated at 250 Hz using Pistonphone Type 4228 with ¼-inch Adaptor DP-0775. The pressure-field response can be measured using Actuator UA- 0033 with Adaptor DB-0264. The pressure-field response is equal to the actuator response.