Types 4227 and Type 4227-A conform to ITU-T Recommendation P.51, delivering 110 dB continuous output and over 120 dB pulsed output for precise, repeatable testing.
Mouth Simulators Type 4227 and Type 4227-A are high-performance artificial mouths designed for accurate and repeatable electroacoustic measurements.
Conforming to International Telecommunication Union Telecommunication Standardization Sector (ITU-T) Recommendation P.51, they generate a sound field that closely replicates human speech, making them ideal for testing telephone transmitters, close-talk microphones and other acoustic transducers.
Both models deliver a guaranteed minimum continuous sound pressure level of 110 dB at 25 mm from the lip ring, with pulsed operation exceeding 120 dB.
Their low-distortion performance, stable acoustic output and built-in overload protection support reliable measurements in development, quality control and production environments.
Type 4227-A additionally includes an integrated Class D amplifier, simplifying setup and reducing heat generation in compact test installations.
Types 4227 and 4227-A function as precision sound sources for evaluating microphones, telephone transmitters and other acoustic devices.
A high-quality loudspeaker mounted within a specially designed housing generates a controlled sound field that closely reproduces the acoustic characteristics of the human mouth.
The integrated lip ring establishes a defined reference plane, ensuring repeatable positioning and measurement conditions.
For applications requiring extremely stable output and a flat frequency response, a reference microphone can be installed in the mouth opening and used within a control loop to maintain constant sound pressure levels.
The simulators are supplied with individual calibration data and a calibration jig to support accurate, traceable measurements.
The mouth simulators integrate easily into laboratory and production test systems.
Type 4227-A features a built-in Class D amplifier that reduces external hardware requirements and minimizes thermal output, making it well suited for installation in compact test chambers.
Both models support integration with reference microphones for closed-loop output control and can be calibrated using the supplied calibration jig with either free-field or pressure-field measurement microphones.
They are compatible with HBK electroacoustic test platforms, including PULSE™ Electroacoustics systems, and can be combined with Telephone Test Head Type 4602 for standardized handset and communication device testing.
Artificial Mouth: A standardized acoustic sound source designed to reproduce the sound field generated by the human mouth, enabling accurate and repeatable testing of microphones, telephones and communication devices.
ITU-T Recommendation P.51: An international standard defining the characteristics and performance requirements of artificial mouths used for testing communication equipment and electroacoustic devices.
Class D Amplifier: A highly efficient amplifier design that generates less heat than conventional amplifiers, making it suitable for compact test systems and continuous operation.
Sound Pressure Level (SPL): A measurement of sound intensity expressed in decibels (dB), commonly used to characterize acoustic output and performance.
Close-talk Microphone: A microphone intended to operate close to the user's mouth, commonly used in communication headsets, telephones and broadcast applications.
| Document | Type | Language |
|---|---|---|
| Mouth Simulator Types 4227 and 4227-A | Data Sheets | English |