TYPE 4178

1/4" 麦克风风对,分机。极化

声强麦克风,专为4 Hz至100 kHz频率范围内的高电平和高频测量而设计, 具有相位响应的宽带匹配。

img

TYPE 4178

expand_more
chevron_left
chevron_right

Type 4178 由一对选定的 TYPE 4939 1/4 英寸麦克风组成,其相位响应在 100 Hz 和 20 kHz 之间匹配。

Ul.
  • 使用双麦克风技术测量声音强度,遵循IEC 61043 Class 1
  • 使用声强探头进行符合ISO 9614-1、ISO 9614-2、ECMA 160和ANSI S12.12标准的声功率测量
  • 声源粒子速度的测量
  • TYPE 4178
    Application Other special microphones

    Features

    • Broadband matching of phase responses
    • Phase calibrated in 1/3-octave frequency bands from 100 Hz – 20 kHz

    Type 4939 microphones are specially designed for high-level and high-frequency measurements. By using stainless steel diaphragm and protection grid, the microphone is optimized to withstand rough environments and is capable of working at high temperatures – up to 150°C (302°F). The microphones are mainly used in Sound Intensity Probes for sound power and intensity measurements, and can also be used for particle velocity measurements.

    Type 4178 is supplied with 6 mm and 12 mm spacers, along with calibration charts giving the individually measured free-field frequency response for each microphone and a Phase Response Comparison Calibration (100 Hz – 20 kHz).

    Calibration

    Phase calibration of 1/4-inch microphone pair Type 4178 is done at Brüel & Kjær by subjecting the two microphones to the same sound signal in a pressure coupler. This individual phase calibration can be used to derive the Pressure-Residual Intensity Index for the microphone pair.

    Sound intensity measurement systems containing a sound intensity probe set can be conveniently calibrated using
    Sound Intensity Calibrator Type 3541-A. This permits simultaneous sensitivity adjustment of both channels of the processor (in pressure, particle velocity, or intensity modes) and allows determination of the 20 Hz – 5 kHz Pressure-Residual Intensity Index of the probe and processor combinations.