MAIN MENU

4192 型

½英寸压力场传声器,外部极化

压力场传声器专为在3.15 Hz至20 kHz的频率范围内进行高精度耦合腔测量或ANSI噪声测量而设计。

img

4192 型

expand_more
chevron_left
chevron_right

4192型专为高精度耦合腔测量或ANSI噪声测量而设计。4192型为外部极化,必须与经典的前置放大器一起使用。

  • 精密耦合腔测量
  • 高级 ANSI 声级计
  • 一般随机入射测量
TYPE 4192
Standard IEC 61094
Sound field Pressure field
Polarization voltage 200V
Sensitivity 12.5 mV/Pa
Lower limiting frequency 3.15 Hz
Higher limiting frequency 20 kHz
Inherent noise 19 dB(A)
Upper limiting of dynamic range 162 dB
Diameter 1/2 inch
Venting Side
Minimum Temperature -30 °C
Maximum Temperature 300 °C
Minimum Temperature -22 °F
Maximum Temperature 572 °F

USE OF PRESSURE-FIELD MICROHONES

Pressure-field microphones are optimised to have a flat frequency response in a pressure field. The microphones should be used for making measurements in small, closed couplers or close to hard, reflective surfaces.

Type 4192 is suited for situations where a robust and stable pressure-field microphone is required.

Stability

A press-fitted, stainless-steel diaphragm ensures superior long-term stability and mechanical robustness – Type 4192 will withstand the 1 m drop test of IEC 60068-2-32.

POLARIZATION VOLTAGE

Type 4192 requires an external polarization voltage and must therefore must be used with a classical preamplifier. Externally polarized microphones may be used at higher temperatures without severe changes in sensitivity.

Type 4192 is available in TEDS combinations with the classical Microphone Preamplifier Type 2669. The TEDS microphone is considered one unit and has been sealed in a clean environment. The TEDS is programmed with the loaded sensitivity of the actual cartridge and the data is therefore readily available. The default TEDS template is to IEEE P 1451.4 but TEDS to IEEE 1451.4 is available on request.