½" DIFFUSE-FIELD / RANDOM INCIDENCE MICROPHONE
Externally polarized, diffuse-field microphone designed for random-incidence measurements over 3.15 Hz to 10 kHz, and is approved for noise measurements in accordance with ANSI standards.
TYPE 4943
Type 4943 is optimised for general, random-incidence measurements and for noise measurements in accordance with ANSI standards. Being externally polarized, it must be used with a classical preamplifier
| Diffuse-field ½-inch Microphone Type 4943 Datasheet (bp1876) | Data Sheets | English |
| Standard | ANSI S1.4 |
| Sound field | Diffuse field |
| Polarization voltage | 200V |
| Sensitivity | 50 mV/Pa |
| Lower limiting frequency | 3.15 Hz |
| Higher limiting frequency | 10 kHz |
| Inherent noise | 15.5 dB(A) |
| Upper limiting of dynamic range | 148 dB |
| Diameter | 1/2 inch |
| Venting | Rear |
| Minimum Temperature | -40 °C |
| Maximum Temperature | 150 °C |
| Minimum Temperature | -40 °F |
| Maximum Temperature | 302 °F |
Un micrófono de campo difuso, también llamado micrófono de incidencia aleatoria, está diseñado para tener una respuesta uniforme cuando las señales llegan simultáneamente desde todos los ángulos.
Por lo tanto, no sólo deben utilizarse para mediciones en cámaras de reverberación, sino en todas las situaciones en las que el campo sonoro es difuso o donde varias fuentes contribuyen a la presión sonora en la posición de medición.
Algunos ejemplos prácticos son situaciones en interiores, donde el sonido se refleja en las paredes, los techos y los objetos de la habitación, o mediciones dentro de un automóvil.
El micrófono es capaz de soportar la prueba de caída de 1 metro según la norma IEC 68-2-32.
La sensibilidad se puede calibrar a 250 Hz utilizando el pistófono Tipo 4229 con nuestro adaptador de ½ pulgada DP 0776, o a 1 kHz utilizando el calibrador de sonido Tipo 4231. Para otras frecuencias, recomendamos utilizar el calibrador acústico multifunción Type 4226.
La respuesta del actuador se puede medir utilizando el actuador UA-0033.
La respuesta de incidencia aleatoria se puede obtener añadiendo la corrección de incidencia aleatoria específica del tipo a la respuesta del actuador.
This will bring together HBM, Brüel & Kjær, nCode, ReliaSoft, and Discom brands, helping you innovate faster for a cleaner, healthier, and more productive world.
This will bring together HBM, Brüel & Kjær, nCode, ReliaSoft, and Discom brands, helping you innovate faster for a cleaner, healthier, and more productive world.
This will bring together HBM, Brüel & Kjær, nCode, ReliaSoft, and Discom brands, helping you innovate faster for a cleaner, healthier, and more productive world.
This will bring together HBM, Brüel & Kjær, nCode, ReliaSoft, and Discom brands, helping you innovate faster for a cleaner, healthier, and more productive world.
This will bring together HBM, Brüel & Kjær, nCode, ReliaSoft, and Discom brands, helping you innovate faster for a cleaner, healthier, and more productive world.
This will bring together HBM, Brüel & Kjær, nCode, ReliaSoft, MicroStrain and Discom brands, helping you innovate faster for a cleaner, healthier, and more productive world.